Acu-rite Micro-Line T Manual do Utilizador Página 14

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 13
MICRO-LINE Reference Manual
SETUP
11
Testing the Scales
Follow these steps to confirm that your scales have been installed proper-
ly. This test will confirm the scale’s electrical operation, and will also
check the installation integrity.
Locate a magnetic base on the machine and set the dial indicator
on the scale’s reading head casting. Zero the readout and the
indicator.
Move the axis through the full travel and return the dial to “0.”
The readout should also read 0 (±.0005” for 10µm scales;
±0.0002” for 5µm scales). If it doesn’t, then the scale cable may
be loose, or the scale, mounting bracket, or reading head may
need to be tightened or realigned.
Repeat these steps for each scale.
MLTmanual.QXD 2/12/02 4:10 PM Page 26
Vista de página 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 45 46

Comentários a estes Manuais

Sem comentários